CVE Vulnerabilities

CVE-2014-5432

Improper Authentication

Published: Mar 26, 2019 | Modified: Oct 09, 2019
CVSS 3.x
9.8
CRITICAL
Source:
NVD
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CVSS 2.x
7.5 HIGH
AV:N/AC:L/Au:N/C:P/I:P/A:P
RedHat/V2
RedHat/V3
Ubuntu

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Weakness

When an actor claims to have a given identity, the product does not prove or insufficiently proves that the claim is correct.

Affected Software

Name Vendor Start Version End Version
Sigma_spectrum_infusion_system_firmware Baxter 6.05 (including) 6.05 (including)

Potential Mitigations

References