CVE-2015-3956

Insufficient Verification of Data Authenticity

Published: Mar 25, 2019 | Modified: Oct 09, 2019
CVSS 3.x
9.8
CRITICAL
Source:
NVD
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CVSS 2.x
10 HIGH
AV:N/AC:L/Au:N/C:C/I:C/A:C
RedHat/V2
RedHat/V3
Ubuntu

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Weakness

The product does not sufficiently verify the origin or authenticity of data, in a way that causes it to accept invalid data.

Affected Software

Name Vendor Start Version End Version
Plum_a+_infusion_system_firmware Pifzer * 13.4 (including)