CVE Vulnerabilities

CVE-2016-8375

Published: Feb 13, 2017 | Modified: Mar 16, 2017
CVSS 3.x
4.9
MEDIUM
Source:
NVD
CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:C/C:H/I:N/A:N
CVSS 2.x
1.9 LOW
AV:L/AC:M/Au:N/C:P/I:N/A:N
RedHat/V2
RedHat/V3
Ubuntu

An issue was discovered in Becton, Dickinson and Company (BD) Alaris 8015 Point of Care (PC) unit, Version 9.5 and prior versions, and Version 9.7, and 8000 PC unit. An unauthorized user with physical access to an affected Alaris PC unit may be able to obtain unencrypted wireless network authentication credentials and other sensitive technical data by disassembling the PC unit and accessing the devices flash memory. The Alaris 8015 PC unit, Version 9.7, and the 8000 PC unit store wireless network authentication credentials and other sensitive technical data on internal flash memory. Accessing the internal flash memory of the affected device would require special tools to extract data and carrying out this attack at a healthcare facility would increase the likelihood of detection.

Affected Software

Name Vendor Start Version End Version
Alaris_8015_pc_unit Bd * 9.5 (including)
Alaris_8015_pc_unit Bd 9.7 (including) 9.7 (including)

References