CVE-2018-10634

Cleartext Transmission of Sensitive Information

Published: Aug 13, 2018 | Modified: May 22, 2025
CVSS 3.x
5.3
MEDIUM
Source:
NVD
CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N
CVSS 2.x
2.9 LOW
AV:A/AC:M/Au:N/C:P/I:N/A:N
RedHat/V2
RedHat/V3
Ubuntu

Communications between Medtronic MiniMed MMT pumps and wireless accessories are transmitted in cleartext. A sufficiently skilled attacker could capture these transmissions and extract sensitive information, such as device serial numbers.

Weakness

The product transmits sensitive or security-critical data in cleartext in a communication channel that can be sniffed by unauthorized actors.

Affected Software

Name Vendor Start Version End Version
Minimed_paradigm_revel_mmt-523k_firmware Medtronic - (including) - (including)