CVE Vulnerabilities

CVE-2019-18254

Cleartext Storage of Sensitive Information

Published: Jun 29, 2020 | Modified: Oct 29, 2021
CVSS 3.x
4.6
MEDIUM
Source:
NVD
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
CVSS 2.x
2.1 LOW
AV:L/AC:L/Au:N/C:P/I:N/A:N
RedHat/V2
RedHat/V3
Ubuntu

BIOTRONIK CardioMessenger II, The affected products do not encrypt sensitive information while at rest. An attacker with physical access to the CardioMessenger can disclose medical measurement data and the serial number from the implanted cardiac device the CardioMessenger is paired with.

Weakness

The product stores sensitive information in cleartext within a resource that might be accessible to another control sphere.

Affected Software

Name Vendor Start Version End Version
Cardiomessenger_ii-s_gsm_firmware Biotronik 2.20 (including) 2.20 (including)

Extended Description

Because the information is stored in cleartext (i.e., unencrypted), attackers could potentially read it. Even if the information is encoded in a way that is not human-readable, certain techniques could determine which encoding is being used, then decode the information. When organizations adopt cloud services, it can be easier for attackers to access the data from anywhere on the Internet. In some systems/environments such as cloud, the use of “double encryption” (at both the software and hardware layer) might be required, and the developer might be solely responsible for both layers, instead of shared responsibility with the administrator of the broader system/environment.