CVE Vulnerabilities

CVE-2019-18256

Insufficiently Protected Credentials

Published: Jun 29, 2020 | Modified: Oct 29, 2021
CVSS 3.x
4.6
MEDIUM
Source:
NVD
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
CVSS 2.x
2.1 LOW
AV:L/AC:L/Au:N/C:P/I:N/A:N
RedHat/V2
RedHat/V3
Ubuntu

BIOTRONIK CardioMessenger II, The affected products use individual per-device credentials that are stored in a recoverable format. An attacker with physical access to the CardioMessenger can use these credentials for network authentication and decryption of local data in transit.

Weakness

The product transmits or stores authentication credentials, but it uses an insecure method that is susceptible to unauthorized interception and/or retrieval.

Affected Software

Name Vendor Start Version End Version
Cardiomessenger_ii-s_gsm_firmware Biotronik 2.20 (including) 2.20 (including)