CVE Vulnerabilities

CVE-2020-12035

Improper Authentication

Published: Jun 29, 2020 | Modified: Nov 21, 2024
CVSS 3.x
4.9
MEDIUM
Source:
NVD
CVSS:3.1/AV:P/AC:L/PR:L/UI:N/S:U/C:N/I:H/A:L
CVSS 2.x
3.6 LOW
AV:L/AC:L/Au:N/C:N/I:P/A:P
RedHat/V2
RedHat/V3
Ubuntu

Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration.

Weakness

When an actor claims to have a given identity, the product does not prove or insufficiently proves that the claim is correct.

Affected Software

Name Vendor Start Version End Version
Prismaflex_firmware Baxter * *

Potential Mitigations

References