CVE Vulnerabilities

CVE-2020-12036

Cleartext Transmission of Sensitive Information

Published: Jun 29, 2020 | Modified: Nov 21, 2024
CVSS 3.x
7.5
HIGH
Source:
NVD
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
CVSS 2.x
5 MEDIUM
AV:N/AC:L/Au:N/C:P/I:N/A:N
RedHat/V2
RedHat/V3
Ubuntu

Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.

Weakness

The product transmits sensitive or security-critical data in cleartext in a communication channel that can be sniffed by unauthorized actors.

Affected Software

Name Vendor Start Version End Version
Prismaflex_firmware Baxter * *

Potential Mitigations

References