CVE Vulnerabilities

CVE-2020-27252

Time-of-check Time-of-use (TOCTOU) Race Condition

Published: Dec 14, 2020 | Modified: Dec 15, 2020
CVSS 3.x
8.1
HIGH
Source:
NVD
CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
CVSS 2.x
9.3 HIGH
AV:N/AC:M/Au:N/C:C/I:C/A:C
RedHat/V2
RedHat/V3
Ubuntu

Medtronic MyCareLink Smart 25000 all versions are vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device.

Weakness

The product checks the state of a resource before using that resource, but the resource’s state can change between the check and the use in a way that invalidates the results of the check. This can cause the product to perform invalid actions when the resource is in an unexpected state.

Affected Software

Name Vendor Start Version End Version
Mycarelink_smart_model_25000_firmware Medtronic * *

Potential Mitigations

References