CVE Vulnerabilities

CVE-2020-27260

Improper Neutralization of Special Elements in Output Used by a Downstream Component ('Injection')

Published: Jan 08, 2021 | Modified: Nov 21, 2024
CVSS 3.x
5.3
MEDIUM
Source:
NVD
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:N
CVSS 2.x
2.1 LOW
AV:L/AC:L/Au:N/C:N/I:P/A:N
RedHat/V2
RedHat/V3
Ubuntu
root.io logo minimus.io logo echo.ai logo

Innokas Yhtymä Oy Vital Signs Monitor VC150 prior to Version 1.7.15 HL7 v2.x injection vulnerabilities exist in the affected products that allow physically proximate attackers with a connected barcode reader to inject HL7 v2.x segments into specific HL7 v2.x messages via multiple expected parameters.

Weakness

The product constructs all or part of a command, data structure, or record using externally-influenced input from an upstream component, but it does not neutralize or incorrectly neutralizes special elements that could modify how it is parsed or interpreted when it is sent to a downstream component.

Affected Software

NameVendorStart VersionEnd Version
Vital_signs_monitor_vc150_firmwareInnokasmedical*1.7.15 (excluding)

Potential Mitigations

References