CVE-2023-25931

Unverified Password Change

Published: Mar 01, 2023 | Modified: Nov 21, 2024
CVSS 3.x
6.8
MEDIUM
Source:
NVD
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CVSS 2.x
RedHat/V2
RedHat/V3
Ubuntu

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.

Weakness

When setting a new password for a user, the product does not require knowledge of the original password, or using another form of authentication.

Affected Software

Name Vendor Start Version End Version
Interstim_x_clinician Medtronic a51300 (including) a51300 (including)
Micro_clinician Medtronic a51200 (including) a51200 (including)

Potential Mitigations